SUMMARY A scanning tunnelling microscope has been designed which allows tunnelling microscopy to be performed in the presence of an externally applied electrochemical current. Separate, isolated electrodes were used for electrochemical control, and up to 1 mA was passed during real‐time, video‐rate, in situ STM observation of the surfaces, without interfering with the operation of the STM. The noise level of these STM images is only slightly higher than images taken with the electrochemical circuitry disconnected. Surfaces were observed during the formation of surface films in aqueous electrolytes.