Microfluidic devices with integrated minisolenoids, microvalves, and channels have been fabricated for fast and low-volume immunoassay using superparamagnetic beads and well-known surface bioengineering protocols. A magnetic reaction area can be formed in the microchannel, featuring a high surface-to-volume ratio and low diffusion distances for the reagents to the bead surface. Such a method has the obvious advantage of easy implementation at low cost. Moreover, the minisolenoids can be switched on or off and the magnetic field intensity can be tuned on demand. Fluids can be manipulated by controlling the integrated air-pressure-actuated microvalves. Accordingly, magnetic bead–based immunoassay, as a typical example of biochemical detection and analysis, has been successfully performed on the integrated microfluidic device automatically in longitudinal mode. With a sample consumption of 0.5μl and a total assay time of less than 15min, goat immunoglobulin G was detected and the method exhibited a detection limit of 4.7ng∕ml.