Answers are grounded in this paper and cite their sources.
We present a portable optically pumped magnetometer instrument for ultrasensitive measurements within the Earth's magnetic field. The central part of the system is a sensor head operating a MEMS-based Cs vapor cell in the light-shift dispersed Mz mode. It is connected to a compact, battery-driven electronics module by a flexible cable. We briefly review the working principles of the device and detail the realization of both the sensor head and electronics. We show shielded and unshielded measurements within a static magnetic field amplitude of 500.2em0exμT, demonstrating a noise level of the sensor system down to 1400.2em0exfT/√Hz and a sensor bandwidth of several 100 Hz. In a detailed analysis of sensor noise, we reveal the system to be limited by technical sources with straightforward strategies for further improvement toward its fundamental noise limit of 120.2em0exfT/√Hz. We assess the parameters defining the sensor bandwidth by theoretical modeling based on the Bloch equations. Finally, we compare our sensors' performance to a commercial superconducting quantum interference device system in a measurement environment typical for geomagnetic observatory practice and geomagnetic prospection.
No takes yet. Share an insight, caveat, or question.
Oelsner et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: