Methods for local photon path length and optical properties estimation, based on measured and simulated diffuse reflectance within 2 mm from the light source, are proposed and evaluated in vivo on Caucasian human skin. The accuracy of the methods was good (2%-7%) for path length and reduced scattering but poor for absorption estimation. Reduced scattering and absorption were systematically lower in the fingertip than in the forearm skin (633 nm). A maximum intrasite and interindividual variation of approximately 35% in an average photon path length was found. The methodology was applied in laser Doppler flowmetry, where path-length normalization of the estimated perfusion removed the optical property dependency.
No takes yet. Share an insight, caveat, or question.
Larsson et al. (2003) studied this question.