Key result
Lower revolution speed during rotational atherectomy (<150,000 rpm) was significantly associated with a greater burr size-corrected debulking area in severely calcified coronary lesions (p=0.018).
Why the study?
The number of patients requiring calcium ablation is increasing, but little evidence exists regarding the association between debulking area and revolution speed during rotational atherectomy.
Does lower revolution speed during rotational atherectomy improve the calcium debulking area in patients with severe calcified coronary lesions?
Cohort (n=29)
Does lower revolution speed during rotational atherectomy improve the calcium debulking area in patients with severe calcified coronary lesions?
Rotational atherectomy at lower revolution speeds (<150,000 rpm) is associated with a greater calcium debulking effect in severely calcified coronary lesions.
Answers are grounded in this paper and cite their sources.
Lower RA speeds <150,000 rpm were associated with greater calcium debulking; leaves open whether this improves procedural or clinical outcomes.
Objectives In this study, we sought to investigate the association between revolution speed of rotational atherectomy (RA) and debulking area assessed by frequency domain‐optical coherence tomography (FD‐OCT). Background The number of patients with severe calcified coronary artery disease requiring treatment with calcium ablation, such as RA, is increasing. However, there is little evidence available regarding the association between debulking area and revolution speed during RA. Methods We retrospectively investigated 30 consecutive severely calcified coronary lesions in 29 patients who underwent RA under FD‐OCT guidance. The association between preset revolution speed of RA and burr size‐corrected debulking area of the calcified lesion was evaluated using a multivariable regression model with nonlinear restricted‐cubic‐spline, which can help assess nonlinear associations between variables. Results The median age of study participants was 73 years (quartile 65–78); 82.8% were male. The median burr size was 1.5 mm (1.5–1.75); median total duration of ablation was 120 s (100–180). FD‐OCT revealed that the post‐procedural minimum lumen area increased significantly from 1.64 mm 2 (1.40–2.09) to 2.45 mm 2 (2.11–2.98) ( p < .001). In addition, the burr size‐corrected debulking area increased significantly as the preset revolution speed decreased ( p = .018), especially when the revolution speed was less than 150,000 rpm. This result implies that additional lumen gain will be obtained by decreasing rpm when the burr speed is set at <150,000 rpm. Conclusions FD‐OCT demonstrated that RA with lower revolution speed, below 150,000 rpm, has the potential to achieve greater calcium debulking effect in patients with severe calcified coronary lesions.
No takes yet. Share an insight, caveat, or question.
Mizutani et al. (2019) conducted a cohort in severe calcified coronary artery disease (n=29). Revolution speed of rotational atherectomy was evaluated on burr size-corrected debulking area (p=0.018). Lower revolution speed during rotational atherectomy (<150,000 rpm) was significantly associated with a greater burr size-corrected debulking area in severely calcified coronary lesions (p=0.018).