BACKGROUND: This study evaluated the clinical accuracy and analytical sensitivity of the NanoSign® Influenza A/B antigen kit in detecting 2009 pandemic influenza A/H1N1 viruses. The kit is one of the most popular rapid diagnostic tests for detecting influenza in Republic of Korea. RESULTS: The NanoSign® Influenza A/B kit resulted in 79.4% sensitivity and 97.2% specificity compared to RT-PCR in the detection of the viruses from 1,023 specimens. In addition, the kit was able to detect two strains of novel influenza viruses, Influenza A/California/12/2009(H1N1) and clinically isolated wild-type novel influenza A/H1N1, both of which are spreading epidemically throughout the world. In addition, the correlation between NanoSign® Influenza A/B test and conventional RT-PCR was approximately 94%, indicating a high concordance rate. Analytical sensitivity of the kit was approximately 73 ± 3.65 ng/mL of the purified viral proteins and 1.13 ± 0.11 hemagglutination units for the cultured virus. CONCLUSIONS: As the NanoSign® Influenza A/B kit showed relatively high sensitivity and specificity and the good correlation with RT-PCR, it will be very useful in the early control of influenza infection and in helping physicians in making early treatment decisions.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2010) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: