Answers are grounded in this paper and cite their sources.
Efficiency and resolution measurements are reported of an x-ray transmission grating to be flown on board the European X-ray Observatory Satellite (EXOSAT). By optimizing the thickness of the grating bars, very high efficiencies can be obtained over a limited energy range. An absolute efficiency of 20% at 7.1 A decreasing to 10% above 15 A has been achieved, including all effects of support structure and mounting necessary to withstand space qualification. This means that 10% of the incoming radiation at 7.1 A results in the minus first spectral order and 10% in the plus first order. The measured resolution at the short wavelength end of the range (lambda > 20 A) is determined by the mirror resolution; at the longer wavelengths (300 A), the effects of coma and astigmatism dominate. The measured resolution at 304 A for the EXOSAT-mirror-grating system is 4.5 A in agreement with theoretical predictions.
No takes yet. Share an insight, caveat, or question.
Brinkman et al. (1980) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: