The absolute co un ting of beta particles, using 4 7r proportional flow counters, is described . This method of absolute co untin g has bee n adopted in the Radioactivity Section of the National Bureau of Standards for the primary calibration of radioisotope solut ion standards. Pillbox, cylindrical , and spherical 4 7r flow counters have been used in the m ea surements, and exce ll ent results have been obtained for each. A simple method of source mounting has bee n developed , and a formula is derived t hat determin es the absorption in the sou rce moun t ing r egardless of knowledge of the t hi ckness of the moun ting or energy spectrum of t he beta rays. R es ults are given of in tercompari son s of p 32 and J131 with other laboratories in the United States, Canada, and Great Bri tain that e mp loy different types of absolute counting techniques. The applicability of the 4 7r countin g method to t he standard ization of practically all beta emit t ing radiosotopes is di scussed .