We performed a randomised, crossover, single-blind trial among 168 patients, to compare the single-use SoftSeal and LMA Unique laryngeal mask airways in spontaneously breathing adults. Size-3 and -4 laryngeal mask airways, inserted by experienced anaesthetists, performed equivalently for successful first-time placement (148 (91%) vs 155 (96%) for the SoftSeal and LMA Unique, respectively). The SoftSeal was more often rated as difficult to insert (27 (17%) vs 4 (3%); p < 0.001) and was more likely to show evidence of mucosal trauma after the first insertion (14 (10%) vs 5 (4%); OR 1.3 (95% CI 1.3-11.3); p < 0.05). The fibreoptic view of the larynx was better through the SoftSeal (vocal cords not visible in 27 (17%) vs 44 (27%); p < 0.05) and it more frequently provided a ventilation seal at 20 cmH(2)O (93 (59%) vs 62 (39%); OR 2.15 (CI 1.44-3.21); p < 0.001). In contrast to the LMA Unique(trade mark), its cuff pressure did not increase during nitrous oxide anaesthesia (median (IQR [range]) decrease 3 (- 20-23 [-40-94]) cmH(2)O vs increase of 16 (-2-39 [-54-112]) cmH(2)O; p < 0.01). Both devices were equivalent for the success of first-time insertion and performed satisfactorily clinically. There were some performance differences, but either appears suitable for airway management in spontaneously breathing patients.
No takes yet. Share an insight, caveat, or question.
Paech et al. (2005) studied this question.