Answers are grounded in this paper and cite their sources.
To the Editor: Hagberg et al. (1) recently reported successful insertion rates of 92.4% on first attempt and 100% on second attempt for the new AMBU Laryngeal Mask. We think that the 90 degree angle incorporated into the airway tube contributes significantly to this high insertion success rate. As previously described by Yodfat (2), creating a 90 degree angle with a rigid stylet close to the laryngeal portion of the laryngeal mask airway (LMA) improves the rate of successful insertion. Jaffe and Brock-Utne (3) described a similar method for the LMA insertion using a stylet folded in half and also bent nearly 90 degrees. These data suggest that a 90 degrees angle close to the junction of the airway tube and mask, either by design, as in the Ambu Laryngeal Mask, or by inserting a rigid stylet, simplifies the insertion technique and improves the rate of successful insertion. Sonia J. Vaida, MD Visiting International Associate Professor Department of Anesthesiology Penn State Milton S. Hershey Medical Center Hershey, Pennsylvania [email protected] [email protected] Uriel A.