Automated platelet counting in concentrates is subject to instrument bias and activated platelet components, requiring correction factors against reference methods.
Answers are grounded in this paper and cite their sources.
Automated platelet counts in concentrates may require instrument-specific corrections; leaves open optimal standardization across platforms.
BACKGROUND AND OBJECTIVES: The blood platelet content (in numbers) of platelet concentrates is required for production quality control and to predict clinical responses. MATERIALS AND METHODS: This study compared the performance of automated counting from impedance and optical instruments to data from immunoplatelet reference analysis. RESULTS: All methods showed good linearity with evidence of significant instrument-specific deviations from the line of agreement. Relational formulae largely corrected bias, but did not resolve platelet count variability. A second confounding factor, related to the proportion of small (activated) platelets, was also shown to contribute to intermethod discrepancies. CONCLUSIONS: Blood processing centres should establish correction factors for each instrument compared to reference methods, such as the immunoplatelet count.