The potentiometric response behavior of Ca2+‐selective poly(vinyl chloride) (PVC) and polyurethane (PU) membranes with different inner contacts has been compared. Evidence for the formation of a water film between membrane and internal contact and, hence, a less than optimal lower detection limit have been found for membranes directly in contact with bare Au or with polypyrrole generated by electropolymerization in the presence of KCl. A significantly better behavior is shown by membranes with polypyrrole prepared in the presence of potassium hexacyanoferrate. Best performances have been obtained with solvent‐cast poly(3‐octylthiophene) as the internal contacting layer. As compared with the PVC membranes, those with PU had a significantly worse performance throughout.