Effect of instrumental factors on the observed intensity distribution of x-ray lines and at absorption limits.---(1) Slit width. In the case of two slits of equal width, sharp lines reflected from a perfect crystal should give an isosceles triangle distribution. If the two slit widths are unequal, the triangle is truncated; this may shift the relative position of the lines of a partly resolved doublet. (2) Slit height widens the isosceles triangles slightly, making the long-wave-length side less steep. (3) Non-uniform energy distribution over the focal spot produces corresponding lack of symmetry. An actual distribution curve obtained with very narrow slits showed two unequal maxima near the edges of the spot. The same three factors affect the apparent sharpness of the absorption limit. In particular, the finite slit width results in the rounding off of a sharp discontinuity into a skew-symmetrical S curve consisting of parts of two parabolas whose axes are vertical and separated a distance equal to the angular slit width.Intensity distribution in the Kα doublet of Mo.---The curves correspond to monochromatic lines of width.0003A or less. The relative intensity is very close to 2 (2.01). There is no line between them with intensity greater than 1 percent that of Kα₁.Sharpness of the K absorption limit of Ag.---The curve obtained agrees with that for a limit which is sharp within experimental error,.0002A.
No takes yet. Share an insight, caveat, or question.
F. K. Richtmyer (1925) studied this question.