This is the authors' abstract. We don't add key points for this paper.
Using a simultaneous phase sensor, the proposed instrument performs highly repeatable measurements over an extended range in the presence of vibration common to a laboratory setting. Measurement of a 4.5 μm step standard in the presence of vibration amplitudes of 40 nm produces a repeatability of 1.5 nm RMS with vertical scanning data acquired at 400 nm intervals. The outlined method demonstrates the potential to tolerate larger vibration amplitudes up to or beyond a quarter wavelength and to increase the data acquisition step size to that approaching the depth of field of standard microscope imaging systems.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Wiersma et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: