Answers are grounded in this paper and cite their sources.
METHODS: Patients treated with metal-on-metal total hip arthroplasty (MoM-THAs) from 2007 to 2012 were preliminarily included. Clinical registered data included: age, gender, BMI, time between surgery and blood sample, physical activity, Harris Hip Score at follow-up, leg length discrepancy and femoral head size. At postoperative X-ray were collected: acetabular inclination and anteversion, femoral offset, stem subsidence and osteolysis. A multivariate linear regression model was used to determine factors that influence metal ion release.RESULTS: A total of 191 MOM-THAs were included. The average levels of metal ions were 2.03 and 1.95 μg/L respectively for cobalt (Co) and chromium (Cr). Thirty-four patients showed values of Cr or Co out of the range of normality. Ten patients (10/34) underwent revision surgery performed in 7 cases for a pseudotumor. The female gender, the UCLA Score and the cup inclination showed a direct correlation with Cr serum levels. The BMI<25 and an increase in femoral offset (FO) showed a direct correlation with Co serum levels.CONCLUSIONS: Our data suggest that female gender, physical activity, cup inclination, BMI<25 and an increase in FO are predictors of higher serum metal ion levels. These patients might be at increased risk for adverse reaction to metal debris thus a particular surveillance is suggested.
No takes yet. Share an insight, caveat, or question.
Schiavi et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: