Key result
Radial tonometric continuous blood pressure monitoring with the Colin Pilot was associated with fewer artifacts compared to the Finapres monitor during reactivity testing.
Why the study?
Does radial tonometric continuous blood pressure monitoring (Colin Pilot) provide accurate measurements with fewer artifacts compared to Finapres and Dinamap during cardiovascular reactivity testing?
Cross-Sectional (n=30)
Does radial tonometric continuous blood pressure monitoring (Colin Pilot) provide accurate measurements with fewer artifacts compared to Finapres and Dinamap during cardiovascular reactivity testing?
Radial tonometric continuous blood pressure monitoring with the Colin Pilot provides highly accurate measurements with low artifact rates during cardiovascular reactivity testing, outperforming the Finapres device.
Answers are grounded in this paper and cite their sources.
May reduce artifacts in reactivity testing; leaves open prospective validation versus established monitors.
BACKGROUND: Non-invasive continuous blood pressure monitoring has, for the past decade, been performed using the Finapres monitor. Radial tonometry is a new methodology for measuring non-invasive beat-to-beat blood pressure. This technique has been evaluated for other applications, but its use in cardiovascular reactivity testing has not yet been reported. DESIGN: A convenience sample was used to compare a radial tonometric device (Colin Pilot) with intermittent brachial blood pressure measurement (Critikon Dinamap) during reactivity testing. A second comparison was made between an Ohmeda Finapres and a Dinamap. METHODS: Twenty subject-testing sessions during a reactivity protocol were evaluated for blood pressure determination. Separate groups of subjects were used for Pilot versus Dinamap comparisons and Finapres versus Dinamap comparisons. Blood pressure signals from 15 subjects with Pilot and 15 subjects with Finapres data were evaluated for artifacts. Signals were also compared in a subject with Raynaud's phenomenon. RESULTS: Both devices tracked blood pressure well. The Finapres significantly overestimated systolic blood pressure ( P < 0.001). The correlations between the Pilot and Dinamap were consistently higher than those between the Finapres and Dinamap. There were more artifacts (3.2%) with the Finapres than with the Pilot (0.2%). In the subject with Raynaud's phenomenon, the Finapres overestimated blood pressure by more than 40 mmHg and showed more than 75% of artifacts. On the other hand, the Pilot tracked blood pressure in the same subject without error. CONCLUSIONS: Radial tonometric continuous beat-to-beat blood pressure monitoring with the Colin Pilot was associated with a low artifact rating and high accuracy of blood pressure measurement during reactivity testing.
No takes yet. Share an insight, caveat, or question.
Nelesen et al. (2002) reported a cross-sectional. Radial tonometric device (Colin Pilot) vs. Ohmeda Finapres and Critikon Dinamap was evaluated on Artifacts during blood pressure monitoring. Radial tonometric continuous blood pressure monitoring with the Colin Pilot was associated with fewer artifacts compared to the Finapres monitor during reactivity testing.