Answers are grounded in this paper and cite their sources.
We developed and validated a liquid chromatography technique to determine organic impurities of Pitolisant in Pitolisant film-coated tablets (4.45 mg and 17.8 mg), a medication used to treat excessive daytime drowsiness. The method uses a gradient elution process on an Inertsil ODS-3 V column (250 mm × 4.6 mm, 3 μm), with a flow rate of 0.8 mL/min and detects substances at 210 nm. The analytical parameters consist of a sample injection volume of 10 μL, a sample temperature of 10 °C, and an analytical column compartment temperature of 55 ° C. mobile phase A is made up of 70 mM sodium perchlorate monohydrate at a pH of 2.0 mixed with acetonitrile in a 75: 25 ratio, while mobile phase B is a mix of methanol and water in a 90:10 ratio. The recovery rate of impurities ranges from 98.9 % to 104.9 %. All impurities exhibited excellent linearity, with correlation values (r 2 ) of 0.9987 or 0.9999. The method's stability-indicating nature was evaluated through forced degradation studies. Its robustness was assessed using a full factorial design, and we identified the method's operable design space.
No takes yet. Share an insight, caveat, or question.
Raghupathi et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: