Key result
Deactivation of the pacemaker and urgent repositioning of leads resulted in immediate cessation of abdominal wall pulsations and lower limb jerking caused by Reel syndrome.
Why the study?
Pacemaker lead dislodgement from device manipulation is a rare complication of permanent pacemaker insertion, and lead stimulation of surrounding structures can cause misleading presentations.
Case Report (n=1)
Pacemaker lead dislodgement (Reel syndrome) can present with non-cardiac symptoms such as abdominal wall pulsations and lower limb jerking, requiring prompt device deactivation and lead repositioning.
Abdominal pulsations may indicate lead dislodgement in paced patients; leaves open whether this represents an underrecognized ratchet syndrome variant.
BACKGROUND: Pacemaker lead dislodgement and failure, related to device manipulation, is a rare complication of permanent pacemaker (PPM) insertion. Reel's, Twiddler's, and Ratchet syndrome are rare causes of pacemaker failure with varying mechanisms, defined by their classical lead and generator findings on chest X-ray imaging. Misleading patient presentations may be attributed to lead stimulation of surrounding structures. CASE SUMMARY: A 77-year-old female was admitted with abdominal wall pulsations, abdominal pain, and lower limb jerking 3 months following PPM insertion. Following exclusion of a ruptured abdominal aortic aneurysm, the presence of Reel syndrome was noted on the patient's chest X-ray and the electrocardiogram showed inappropriate pacing. Deactivation of the pacemaker resulted in immediate symptom cessation and urgent repositioning of pacemaker leads was undertaken. DISCUSSION: This case highlights the importance of considering pacemaker complications causing non-cardiac symptomatology. Pacemaker lead stimulation of surrounding structures can present in an unconventional fashion, veiling the diagnosis. However, a structured approach to undifferentiated neuromuscular presentations in patients with PPMs should consider lead dislodgement as a differential diagnosis. Rapid recognition of lead dislodgement, device deactivation, and re-implantation or repositioning of the leads are critical in preventing potentially life-threatening complications.
No takes yet. Share an insight, caveat, or question.
Bellinge et al. (2021) conducted a case report in Reel syndrome (pacemaker lead dislodgement) (n=1). Pacemaker deactivation and lead repositioning was evaluated on Symptom cessation. Deactivation of the pacemaker and urgent repositioning of leads resulted in immediate cessation of abdominal wall pulsations and lower limb jerking caused by Reel syndrome.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: