Key result
Pacemaker reel syndrome causes complete heart block and total device failure in an 84-year-old patient.
Case Report (n=1)
Highlights the risk of pacemaker failure due to reel syndrome (device manipulation) in elderly patients with cognitive impairment.
Answers are grounded in this paper and cite their sources.
Reel syndrome risks late pacemaker failure in cognitively impaired elderly; leaves open optimal prevention strategies.
Twiddler's syndrome is a rare cause of pacemaker failure, where patient manipulation of the pulse generator results in lead dislodgement or retraction. Variations in manifestation have been identified including reel syndrome, where rotation occurs around the transverse axis resulting in coiling of the leads, and ratchet syndrome where arm movement results in lead displacement. Device manipulation leading to device failure has been documented in up to 1.7% of implants, particularly in patients with large pockets or mental disorders. Such complications have serious consequences, particularly in pacing-dependent patients where loss of capture may result in asystole. This article reviews the case of an 84-year-old patient presenting at 8-month pacemaker follow-up in complete heart block with no evidence of pacemaker function.
No takes yet. Share an insight, caveat, or question.
Smith et al. (2020) conducted a case report in Pacemaker reel syndrome (n=1). Pacemaker was evaluated. Pacemaker reel syndrome caused complete heart block and total loss of device function in an 84-year-old patient at 8-month follow-up.