Key result
The McGRATH MAC size 1 blade reduced the median time to tracheal intubation compared to the size 2 blade in infants undergoing cardiac surgery.
Why the study?
To determine whether the recently available McGRATHTM MAC size 1 blade reduces tracheal intubation time in infants compared with the size 2 blade.
Does the McGRATH MAC size 1 blade reduce tracheal intubation time in infants undergoing cardiac surgery compared to the size 2 blade?
Observational (n=68)
No
Does the McGRATH MAC size 1 blade reduce tracheal intubation time in infants undergoing cardiac surgery compared to the size 2 blade?
Using the McGRATH MAC size 1 blade for tracheal intubation in infants undergoing cardiac surgery significantly reduces intubation time compared to the size 2 blade.
Answers are grounded in this paper and cite their sources.
May support size 1 blade preference for infant intubation; leaves open confirmation via randomized trials in cardiac surgery.
Purpose: This study aimed to determine whether the McGRATHTM MAC size 1 blade, which has become available in recent years, reduced the tracheal intubation time in infants compared to that of the McGRATHTM MAC size 2 blade. Methods: This single-center, retrospective, observational study included patients aged < 1 y who underwent cardiac surgery under general anesthesia. Cases in which the McGRATHTM MAC size 2 blade was used were classified as MAC2 and those in which the McGRATHTM MAC size 1 blade was used as MAC1. The primary endpoint was the difference in time required for tracheal intubation. The results were adjusted for confounding factors (patient weight, career of intubation provider, and risk of difficult tracheal intubation). Other variables included the first successful tracheal intubation rate, number of tracheal intubations, Cormack classification, and adverse events during tracheal intubation. Results: In total, 94 patients met the inclusion criteria, 26 met the exclusion criteria¸ thus 68 patients were analyzed. The median time to tracheal intubation was significantly shorter for MAC1 than for MAC2. (23.5 s vs. 27.0 s; hazard ratio, 1.79; 95% confidence interval, 1.08, 2.95). The adjusted time to tracheal intubation was also significantly shorter in the MAC1 group than in the MAC2 group (hazard ratio, 1.71; 95% confidence interval, 1.00, 2.89). No secondary endpoint showed significant differences. Conclusions: For tracheal intubation in cardiac surgery in infants, using the McGRATHTM MAC size 1 blade may reduce the tracheal intubation time compared to the McGRATHTM MAC size 2 blade.
No takes yet. Share an insight, caveat, or question.
Uchinami et al. (2023) conducted an observational in Cardiac surgery (n=68). McGRATH MAC size 1 blade vs. McGRATH MAC size 2 blade was evaluated on Time required for tracheal intubation (seconds) (HR 1.79, 95% CI 1.08, 2.95). The McGRATH MAC size 1 blade reduced the median time to tracheal intubation compared to the size 2 blade in infants undergoing cardiac surgery.