Key result
A review of 30 studies evaluating 23 devices found that the accuracy of automated blood pressure measuring devices appears to decrease at increasing blood pressure levels.
Why the study?
Does the accuracy of automated blood pressure measuring devices decrease at increasing blood pressure levels?
Systematic Review (n=30)
Does the accuracy of automated blood pressure measuring devices decrease at increasing blood pressure levels?
Automated blood pressure measurement devices appear less accurate at higher blood pressure levels, though this may be an artifact of increased blood pressure variability and sequential measurement protocols.
Answers are grounded in this paper and cite their sources.
Verify high automated BP readings manually; challenges device validation protocols that overlook BP-dependent accuracy.
BACKGROUND: In validation studies reporting on the accuracy of blood pressure measuring devices (ambulatory and non-ambulatory systems), it is frequently stated that the accuracy of blood pressure devices seems to decrease at increasing blood pressure levels. This has been shown for several ambulatory devices in the past. Whether more recently validated devices are less accurate at increasing blood pressure levels is unknown, however. OBJECTIVES: We therefore retrospectively searched the literature for studies performed between 1993 and 2003, reporting on the accuracy of blood pressure measuring devices over different blood pressure levels. When needed, additional information from the authors was requested. METHODS: In total, 30 studies were selected. Of these, the studies reporting on the accuracy of 14 different ambulatory and nine different non-ambulatory devices were useful. For both ambulatory and non-ambulatory devices, accuracy appeared to decrease at increasing blood pressure levels. This was particularly shown for systolic blood pressure. RESULTS: We speculate whether this finding is due to the oscillometric method of blood pressure measurement. Another explanation may exist, however. Blood pressure variability increases with higher blood pressure. Further, the British Hypertension Society protocol 1993 uses sequential measurements. This may be the reason that, owing to the increased blood pressure variability, the accuracy of most devices tends to decrease at higher blood pressure levels. Consequently, the accuracy of blood pressure measuring devices may be underestimated at higher blood pressure levels. CONCLUSION: Currently used automated blood pressure measurement devices seem to be less accurate at increasing blood pressure levels. It is important to be aware of this phenomenon when treating hypertensive patients. The reported decrease in accuracy, however, may well be explained by the increasing blood pressure variability at increasing blood pressure and the use of sequential measurements. If this is the case, then the accuracy of these devices is perhaps underestimated.
No takes yet. Share an insight, caveat, or question.
Braam et al. (2005) conducted a systematic review in Hypertension (n=30). Automated blood pressure measurement devices was evaluated on Accuracy at increasing blood pressure levels. A review of 30 studies evaluating 23 devices found that the accuracy of automated blood pressure measuring devices appears to decrease at increasing blood pressure levels.