Key result
Double-pulse defibrillation with 85% FCL separation achieved defibrillation using the same total energy as a single shock in single-pathway lead configurations, reducing shock currents by 20% to 30%.
Why the study?
Does double-pulse defibrillation based on fibrillation cycle length reduce the energy required for successful defibrillation compared to single-shock defibrillation in anesthetized dogs?
Does double-pulse defibrillation based on fibrillation cycle length reduce the energy required for successful defibrillation compared to single-shock defibrillation in anesthetized dogs?
Double-pulse defibrillation with 85% FCL separation can achieve successful defibrillation using the same total energy as a single shock, reducing peak currents by 20-30%.
INTRODUCTION: We investigated a method of defibrillation in which two shocks were delivered to the same electrodes with a separation based on the cycle length of the fibrillation event (FCL). METHODS AND RESULTS: In pentobarbital anesthetized dogs, a computerized system determined the FCL from the fibrillation event, computed the desired double-pulse (DP) shock separation, and immediately delivered the DP shocks. In group 1, energy for 50% success at defibrillation (E50) was measured using separations from 55% to 95% of the FCL and remeasured after administration of flecainide, clofilium, or vehicle to change the FCL. Both drugs increased FCL by approximately 25%. Plots of E50 versus %FCL aligned before and after drug showed that the optimum pulse separation followed the FCL. In group 2, E50s were measured for 55% to 185% FCL separations in clofilium or vehicle-treated animals. The optimum DP E50 was at 85% FCL and was not significantly different from the single-shock E50. In group 3, no differences were found when comparing the probability of success versus total energy relationships for single and optimum DP shocks. Group 4 compared E50s for single and DP shocks using a single-catheter configuration and three-electrode configurations (catheter-subdermal patch). E50s for single and DP shocks were equal using the single-catheter configuration but DP shocks required approximately 20% more energy in the three-electrode configurations. CONCLUSION: In single-pathway lead configurations, two shocks with 85% FCL separation can be reliably combined to defibrillate using the same total energy as a larger single shock. Since the energy is unchanged but the total duration is doubled, DP shock currents are reduced by 20% to 30%.
No takes yet. Share an insight, caveat, or question.
Sweeney et al. (1994) studied Fibrillation. Double-pulse defibrillation vs. Single-shock defibrillation was evaluated on Energy for 50% success at defibrillation (E50). Double-pulse defibrillation with 85% FCL separation achieved defibrillation using the same total energy as a single shock in single-pathway lead configurations, reducing shock currents by 20% to 30%.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: