We developed a kinetic assay for calcium in serum and urine, based on the activation of porcine pancreatic alpha-amylase (EC 3.2.1.1) with 2-chloro-4-nitrophenyl-alpha-maltotrioside as substrate. The kinetic generation of 2-chloro-4-nitrophenol, monitored at 405 nm, was proportional to the concentration of calcium in serum and urine. The assay was developed and evaluated with the Cobas Bio centrifugal analyzer. The average within-run and between-day imprecision (CV) was 0.96/1.26% and 1.07/1.63%, respectively, for serum calcium and 1.50/2.54% and 1.70/2.64%, respectively, for urine calcium. Results of the proposed method (y) correlated well with those determined by atomic absorption spectrophotometry (x): y(serum) = 1.005x + 0.028 mmol/L (Sy/x = 0.058, r = 0.974, n = 50), and y(urine) = 1.017x-0.115 mmol/L Sy/x = 0.30, r = 0.981, n = 25). We also present data showing that the method is highly sensitive, rapid, relatively free of interference, and amenable to automation.