Answers are grounded in this paper and cite their sources.
We have used a resonating mechanical cantilever to detect immunospecific binding of viruses, captured from liquid. As a model virus, we used a nonpathogenic insect baculovirus to test the ability to specifically bind and detect small numbers of virus particles. Arrays of surface micromachined, antibody-coated polycrystalline silicon nanomechanical cantilever beams were used to detect binding from various concentrations of baculoviruses in a buffer solution. Because of their small mass, the 0.5μm×6μm cantilevers have mass sensitivities on the order of 10−19g∕Hz, enabling the detection of an immobilized AcV1 antibody monolayer corresponding to a mass of about 3×10−15g. With these devices, we can detect the mass of single-virus particles bound to the cantilever. Resonant frequency shift resulting from the adsorbed mass of the virus particles distinguished solutions of virus concentrations varying between 105 and 107pfu∕ml. Control experiments using buffer solutions without baculovirus showed small amounts (<50attograms) of nonspecific adsorption to the antibody layer.
No takes yet. Share an insight, caveat, or question.
Ilic et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: