A system for measurement of surface roughness based on total integrated scattering at 532 and 355 nm is built and demonstrated. Surfaces up to 25 mm × 25 mm are scanned in 6 min with a spatial resolution of 0.4 mm. Careful attention to reducing stray light and purging the measurement chamber with filtered air allow scattering resolution better than 10−5. Surface roughness measurements better than 1 nm RMS are demonstrated and confirmed by comparison measurements with an atomic-force microscope.