Answers are grounded in this paper and cite their sources.
RATIONALE: This paper presents an in-depth evaluation of the laser diode thermal desorption (LDTD) device interfaced to a mass spectrometer and identifies the critical method parameters to evaluate when developing a new procedure. METHODS: Samples were solvent extracted and spotted in a 96-well plate. In the case of biological fluids, hydrolysis followed by solid phase extraction is required. The solvent in the 96-well plate is evaporated, followed by mass spectrometric analysis (MS) with atmospheric pressure chemical ionization (APCI). Where applicable, the instrument is operated in a data-dependent mode, with a full-scan mass spectrum followed by MS/MS spectra with a total runtime of 18 s per well. RESULTS: Five key experiments (sample dilution, gas flow, laser ramp, ion saturation/ion suppression, and signal enhancement) were evaluated as critical parameters that could potentially be encountered in method development, and practical solutions are proposed. CONCLUSIONS: Interfacing LDTD with a mass spectrometer allows for rapid screening for many analytes in a wide range of samples, with either minimal or complex sample preparation. Addressing all critical method parameters can aid in creating a sensitive and robust procedure.