Electrochemical sensors based on boron‑doped diamond (BDD) electrodes offer high chemical stability, a wide potential window, and low background current, making them promising candidates for dissolved ozone monitoring. In this study, a compact surface‑mount ozone sensor was developed by integrating chip‑processed BDD electrodes onto printed wiring boards. The sensor exhibited stable voltametric behavior under static water conditions, and linear sweep voltammetry (LSV) showed a strong linear correlation between peak current and ozone concentration determined by UV absorbance. Sensor‑to‑sensor variation was low, demonstrating the reproducibility of the fabrication process. Low‑concentration measurements demonstrated the ability to detect ozone at 0.1 mg/L. Chronoamperometry (CA) provided rapid quantification with a linear response above approximately 0.5 mg/L, enabling substantially shorter measurement times than LSV in the applicable concentration range. These results indicate that the proposed surface‑mount BDD sensor provides a practical platform for compact and rapid ozone concentration monitoring. Further evaluation under lower concentration ranges, diverse water‑quality conditions, and long‑term operation will support future practical deployment.