Answers are grounded in this paper and cite their sources.
An analytical method based on liquid chromatography–tandem mass spectrometry (LC-MS-MS) was developed and validated for the determination of glimepiride in human plasma. Glimepiride and glimepiride-d4 (internal standard) were extracted from the plasma by solid phase extraction and chromatographed on a C18 analytical column using an isocratic mobile phase. Detection was carried out by positive electrospray ionization (ESI+) in multiple reaction monitoring (MRM) mode. The chromatographic separation was obtained within 2 min. Linearity was established from 1.00–500.00 ng/mL with a coefficient of determination (r2) of 0.9981 or greater. The lower limit of quantitation (LOQ) was identifiable and reproducible at 1.00 ng/mL with a precision of 3.8%. The method has shown remarkable reproducibility, with intra- and inter-day precision and accuracy <11.3% and <11.0%, respectively. The method was successfully applied for the quantitation of glimepiride in human plasma to support clinical and pharmacokinetic studies.
No takes yet. Share an insight, caveat, or question.
Dutta et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: