Answers are grounded in this paper and cite their sources.
Biomagnetic measurements have been made since 1963. Since then measurement devices have been developed with increasing sensitivity. When performing biomagnetic investigations, one hs to choose between an expensive high-performance shielding and a cheaper but sophisticated and more noisy gradiometer technique. Combined methods of shielding and gradiometers are difficult because of the distortion of the field homogeneity due to shielding material. Nevertheless, combined means of weak shielding and balanced gradiometers can be satisfactory compromise with adequate signal-to-noise ratio and reasonable expense. In the Ragnar Granit Institute at Tampere University of Technology a shielding chamber consisting of thick aluminium walls has been used. For this study a DS SQUID system with a well-balanced second-order gradiometer was developed to improve the signal-to-noise ratio. The results of this work are presented.