Answers are grounded in this paper and cite their sources.
A reflection-mode photoacoustic (PA) imaging system was designed and built to image deep structures in biological tissues. We chose near-infrared laser pulses of 804-nm wavelength for PA excitation to achieve deep penetration. To minimize unwanted surface signals, we adopted dark-field ring-shaped illumination. This imaging system employing a 5-MHz spherically focused ultrasonic transducer provides penetration up to 38 mm in chicken breast tissue. At the 19-mm depth, the axial resolution is 144 microm and the transverse resolution is 560 microm. Internal organs of small animals were imaged clearly.
No takes yet. Share an insight, caveat, or question.
Song et al. (2007) studied this question.