Answers are grounded in this paper and cite their sources.
A highly sensitive probe based on the skin effect was developed to measure thin film permeability. A new microstrip-line-type probe on a flexible substrate was fabricated and placed in contact with a magnetic thin film. The probe enhanced the signal-to-noise ratio and broadband measurement. The permeability of amorphous CoNbZr film (25 mm × 25 mm, 5 nm thick) and that of CoFeB film (45 mm × 25 mm, 0.5 μm thick) were optimized. The measured values were in rough agreement with theoretical values based on the Landau–Lifshitz–Gilbert equation and eddy current generation up to 40 GHz.
No takes yet. Share an insight, caveat, or question.
Yabukami et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: