Key result
Numerical simulations of the Rotaflow centrifugal blood pump revealed that the Ventricular Assist Device (VAD) condition had the lowest modified index of hemolysis (1170) compared to ECMO (7570) and full-load (9140) conditions.
Population
Computational model of the Rotaflow centrifugal blood pump
Comparison
Operation under ECMO, VAD, and full-load… vs Comparison between the different operational…
Design
Preclinical
Authors
Loading...
Simulations indicate lowest hemolysis under VAD conditions; leaves open clinical translation for Rotaflow pump use.
Numerical simulations of the Rotaflow centrifugal blood pump indicate that hemolysis risk is lowest under VAD conditions and highest under full-load conditions.
Chang et al. (2015) studied this question. Rotaflow centrifugal blood pump vs. Different operational conditions (ECMO, VAD, full-load) was evaluated on Modified index of hemolysis (MIH). Numerical simulations of the Rotaflow centrifugal blood pump revealed that the Ventricular Assist Device (VAD) condition had the lowest modified index of hemolysis (1170) compared to ECMO (7570) and full-load (9140) conditions.