After calibration by thermodilution (TD), the PiCCO device is able to assess cardiac index (CI) using pulse contour (PC) analysis. The manufacturer suggests recalibration by TD after 8 hours. Recently, we suggested a calibration index indicating a certain probability of a relevant bias and triggering the next calibration. With changes of CIpc compared with the previous CItd being a key predictor of the bias of CIpc compared with the following CItd, the manufacturer implemented a soft alarm indicating changes in CIpc compared with the last CItd (thresholds adjustable to 15%, 25% and 35%). The aim of the study was prospective evaluation of predictive capabilities of the 15% alarm regarding a bias of CIpc compared with CItd exceeding critical thresholds of 15%, 20% and 25%.