Key result
Holographic interferometry detected structural irregularities in 8 of 9 porcine bioprostheses, whereas no anomalies were found in 4 fresh porcine aortic valves.
Why the study?
Can holographic interferometry detect hidden defects in bioprosthetic valves in vitro?
Can holographic interferometry detect hidden defects in bioprosthetic valves in vitro?
Absolute Event Rate: 88.9% vs 0%
Holographic interferometry can detect hidden structural defects in bioprosthetic valves in vitro, potentially preventing the implantation of dysfunctional valves.
Dysfunction of heart valve prostheses—mechanical as well as biological—is a common problem in cardiac surgery. The reasons for the valve failures are still not well understood. Biological valves especially have an unsatisfactory durability; degeneration and calcification very often lead to the failure of the valves. In our opinion, hidden defects present in the valve material prior to implantation of the valves is a plausible explanation for the dysfunction. Hitherto there has been no technique to detect these defects without destructing the specimen. Holographic interferometry proved to be applicable for in vitro evaluation of mechanical heart valve prostheses. In the present paper we describe application of this method to biological valves. Nine porcine bioprostheses and four fresh porcine aortic valves were investigated by means of holographic interferometry. In eight of nine bioprostheses, the results showed irregularities of the leaflet structure which depend on anomalies of the connective tissue of the leaflets of the valves. To make sure that these findings are not due to normal variations of the morphology, the investigations were carried out with fresh and unfixated porcine aortic valves. In the latter, no such anomalies of the structure were detected. The results obtained confirm the above hypothesis on the origin of the later valve dysfunction. Thus, holographic interferometry tests of bioprostheses prior to their implantation prevent the use of potentially dysfunctional valves.
No takes yet. Share an insight, caveat, or question.
Geiger et al. (2005) studied Heart valve prosthesis dysfunction (n=13). Holographic interferometry vs. Fresh and unfixated porcine aortic valves was evaluated on Detection of irregularities of the leaflet structure. Holographic interferometry detected structural irregularities in 8 of 9 porcine bioprostheses, whereas no anomalies were found in 4 fresh porcine aortic valves.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: