Key result
In vivo testing of an electrohydraulic actuated total artificial heart in 4 calves demonstrated pump outputs of 4.0-9.5 L/min at left atrial pressures of 7-16 mmHg, suggesting potential for clinical use.
Why the study?
Does an electrohydraulic actuated total artificial heart provide stable hemodynamics in a calf model?
Does an electrohydraulic actuated total artificial heart provide stable hemodynamics in a calf model?
Answers are grounded in this paper and cite their sources.
Initial in vivo testing of an electrohydraulic actuated total artificial heart in calves demonstrated stable hemodynamics, suggesting potential for total implantable cardiac replacement.
The authors are involved in developing a total artificial heart (TAH) for permanent human use. This device was designed to fit human anatomy, and it has housings made of carbon fiber-epoxy composite and titanium. Tissue valves and protein coating of blood contacting surfaces minimize the need for anticoagulants. A continuously reciprocating electrohydraulic actuator is packaged between two alternately ejecting and passively filling ventricles. The control system varies the pump rate to maintain average left ventricular filling at 90%. This TAH in vivo successively progressed through 1, 5, 9, and 45 day implants in calves of 84, 94, 82, and 82 kg preoperative body weights. The operating modes include automatic and fixed rate. The chronic and acute effects of varying the right pump displaced stroke volume indicated the need for it to be limited to 85% of that of the left for stable hemodynamics at maximum flow. The pump exhibited afterload insensitive and preload sensitive performance. Pump output ranged from 4.0-9.5 L/min at left atrial pressures of 7-16 mmHg at pump rates of 80-160 beats/min in these four experiments. These data suggest that this device will meet clinical hemodynamic requirements; it has the potential for total implantable cardiac replacement.
No takes yet. Share an insight, caveat, or question.
IRIÉ et al. (1992) studied Total artificial heart replacement (in vivo testing) (n=4). Electrohydraulic actuated total artificial heart (TAH) was evaluated on Hemodynamic performance (pump output, left atrial pressures, pump rates). In vivo testing of an electrohydraulic actuated total artificial heart in 4 calves demonstrated pump outputs of 4.0-9.5 L/min at left atrial pressures of 7-16 mmHg, suggesting potential for clinical use.