Key result
Infection (20.2%) and device malfunction (20.1%) were the most common adverse events reported following SynCardia total artificial heart implantation, with death occurring in 49.4% of reports.
Why the study?
Real-world clinical data outside of clinical trials and expert centers regarding adverse events related to the SynCardia total artificial heart remain limited.
What are the real-world adverse events associated with the SynCardia total artificial heart reported to the FDA MAUDE database?
Observational (n=453)
What are the real-world adverse events associated with the SynCardia total artificial heart reported to the FDA MAUDE database?
Real-world data from the FDA MAUDE database highlights infection and device malfunction as the most frequent adverse events associated with the SynCardia total artificial heart, emphasizing the need for early recognition and management.
Answers are grounded in this paper and cite their sources.
MAUDE reports identify infection and device malfunction as leading TAH adverse events; leaves open need for prospective studies to inform real-world management.
Background and Objectives: Real-world clinical data, outside of clinical trials and expert centers, on adverse events related to the use of SyncCardia total artificial heart (TAH) remain limited. We aim to analyze adverse events related to the use of SynCardia TAH reported to the Food and Drug Administration (FDA)'s Manufacturers and User Defined Experience (MAUDE) database. Methods: We reviewed the FDA's MAUDE database for any adverse events involving the use of SynCardia TAH from 1/01/2012 to 9/30/2020. All the events were independently reviewed by three physicians. Results: A total of 1,512 adverse events were identified in 453 "injury and death" reports in the MAUDE database. The most common adverse events reported were infection (20.2%) and device malfunction (20.1%). These were followed by bleeding events (16.5%), respiratory failure (10.1%), cerebrovascular accident (CVA)/other neurological dysfunction (8.7%), renal dysfunction (7.5%), hepatic dysfunction (2.2%), thromboembolic events (1.8%), pericardial effusion (1.8%), and hemolysis (1%). Death was reported in 49.4% of all the reported cases (n=224/453). The most common cause of death was multiorgan failure (n=73, 32.6%), followed by CVA/other non-specific neurological dysfunction (n=44, 19.7%), sepsis (n=24, 10.7%), withdrawal of support (n=20, 8.9%), device malfunction (n=11, 4.9%), bleeding (n=7, 3.1%), respiratory failure (n=7, 3.1%), gastrointestinal disorder (n=6, 2.7%), and cardiomyopathy (n=3, 1.3%). Conclusions: Infection was the most common adverse event following the implantation of TAH. Most of the deaths reported were due to multiorgan failure. Early recognition and management of any possible adverse events after the TAH implantation are essential to improve the procedural outcome and patient survival.
No takes yet. Share an insight, caveat, or question.
Tan et al. (2023) conducted an observational in End-stage heart failure (n=453). SynCardia total artificial heart was evaluated on Adverse events and causes of death. Infection (20.2%) and device malfunction (20.1%) were the most common adverse events reported following SynCardia total artificial heart implantation, with death occurring in 49.4% of reports.