Demonstrates 96.7% one-year in vitro reliability for VentrAssist left ventricular assist systems in mock loops, indicating high durability for cardiovascular support.
Left ventricular assist systems (LVADs) are typically tested for long-term reliability by using a simulated human cardiovascular system. In this study, 13VentrAssist (Ventracor Ltd, Chatswood, NSW, Australia)systems are tested in mock loops with the aim of demonstrating in vitro continuous operational reliability of the pump subsystem greater than 80% (60%confidence) over 1 year. As of July 9, 2007, the total time on test of all 13 mock loop pumps is 29.36 years. In vitro reliability of the VentrAssist LVAD blood pump is 96.7%(60% confidence) at 1 year. No failures have been seen over 4.5 years of test, and testing is ongoing.