Key result
An enzyme immunoassay yielded a falsely elevated serum digoxin concentration of 5.0 ng/ml due to marked digoxin-like immunoreactive factor interference, whereas fluorescence polarization immunoassay results were within the therapeutic range.
Case Report (n=1)
Digoxin-like immunoreactive factors in patients with renal insufficiency can cause marked false elevations in serum digoxin levels measured by enzyme immunoassay, necessitating verification with alternative methods.
No takes yet. Share an insight, caveat, or question.
May prompt assay verification for unexpected digoxin levels in renal insufficiency; case report leaves open the prevalence of DLIF interference.
Karboski et al. (1988) conducted a case report in Renal insufficiency (n=1). Enzyme immunoassay vs. Fluorescence polarization immunoassay was evaluated on Serum digoxin concentration (SDC). An enzyme immunoassay yielded a falsely elevated serum digoxin concentration of 5.0 ng/ml due to marked digoxin-like immunoreactive factor interference, whereas fluorescence polarization immunoassay results were within the therapeutic range.