Key result
Long-lasting AC shocks entrain the tissue and decrease defibrillation thresholds at higher frequencies due to stronger virtual electrode polarization compared to single-period waveforms.
Why the study?
Does long-lasting AC shock entrainment lower defibrillation thresholds compared to single-period AC waveforms in a computational canine heart model?
Does long-lasting AC shock entrainment lower defibrillation thresholds compared to single-period AC waveforms in a computational canine heart model?
Long-lasting AC shocks lower defibrillation thresholds at high frequencies by entraining the tissue and creating stronger virtual electrode polarization compared to single-period waveforms.
Preclinical findings on AC waveform optimization warrant no clinical change; leaves open translation to human defibrillation efficacy.
INTRODUCTION: A heart in fibrillation can be entrained by long-lasting alternating current (AC) stimuli, leading to defibrillation. To investigate the role entrainment plays in defibrillation, computer simulations of AC cardioversion in a three-dimensional slice of the canine heart were performed. METHODS AND RESULTS: A bidomain finite element model of a 1-mm thick slice across the ventricles of a canine heart was used to simulate termination of transmural reentry with AC shocks. Cardioversion defibrillation thresholds (DFTs) were determined for 200-msec (L) AC shocks at varying frequencies. At the DFT, the entire tissue is entrained by the AC shock. DFT decreases as the frequency of the long-lasting AC shock increases. We hypothesize that this decrease is due to the short period of the high-frequency AC waveform, leaving strong virtual electrode polarization (VEP) after the shock ends. To test this hypothesis, the end-shock VEP were compared for different frequencies, demonstrating stronger polarization as frequency increased. To examine whether entrainment by the long-lasting AC shock contributes to the VEP at the end of the shock, additional simulations were conducted using single-period (Z) AC waveforms. Z waveform DFTs were higher than L waveform DFTs; the Z waveform VEP was weaker than the L waveform VEP at the same frequency. This indicates that entrainment contributes to the development of stronger VEP and, thus, to lower DFT at high frequencies. CONCLUSION: This study offers for the first time a mechanistic insight into cardioversion with long-lasting AC shocks.
No takes yet. Share an insight, caveat, or question.
Meunier et al. (2002) studied Fibrillation / Transmural reentry. Long-lasting alternating current (AC) shocks vs. Single-period (Z) AC waveforms was evaluated on Cardioversion defibrillation thresholds (DFTs) and virtual electrode polarization (VEP). Long-lasting AC shocks entrain the tissue and decrease defibrillation thresholds at higher frequencies due to stronger virtual electrode polarization compared to single-period waveforms.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: