Authors
We describe a liquid helium temperature regulator of unusually simple design, operation, and construction utilizing miniature solid state operational amplifiers. The combination of a unique saturating amplifier and an analog multiplier type of cross-correlation phase detection circuit eliminates fast risetime signals and switching transients, enabling the regulator to be used in a low noise shielded environment. The circuit is of the conventional ac Wheatstone bridge type using a carbon resistance thermometer for the sensing element and providing a proportional heater feedback of >1 W to the bath. The balanced input circuitry has been optimized for thermometer resistances between 500 Ω and 5 kΩ, and the regulator is capable of detecting a temperature change of 30 μK at a thermometer power dissipation of <3×10−11 W. Regulation of a superfluid 4He bath to better than ±4 μK for periods greater than 30 min has been achieved at a thermometer power of <5×10−9 W, with a response time of 0.1 sec. Owing to the high gain and broad bandwidth of this regulator, it is adaptable to a wide variety of other situations, such as a sample mounted on a copper block outside the bath, without encountering the usual problems of excessive power dissipation, instability, and thermal oscillations.
No takes yet. Share an insight, caveat, or question.
Gene I. Rochlin (1970) studied this question.