An alternative method is presented for calibrating the six-port reflectometer. Through the use of a redundant set of calibration standards, an estimate of the 11 real calibration constants is determined in the minimum-mean-squared-error sense. This technique enables the user to weight the contribution of each standard to the calibration process as a function of confidence in the quality of that standard. The resulting computer algorithm is quite straightforward and provides a direct measure of the tightness of fit between the estimated six-port model and the observed data.>