Key result
A new bradycardia diagnosis algorithm in a dual chamber pacemaker detected at least one bradycardia episode in 54.2% (13 of 24) of patients over a mean follow-up of 153 days.
Why the study?
Does a new bradycardia diagnosis algorithm in a dual chamber pacemaker effectively detect bradycardia in patients considered candidates for diagnostic pacemakers?
Observational (n=24)
Does a new bradycardia diagnosis algorithm in a dual chamber pacemaker effectively detect bradycardia in patients considered candidates for diagnostic pacemakers?
A novel bradycardia diagnosis algorithm in a dual chamber pacemaker is feasible for detecting paroxysmal bradycardia in patients with unexplained syncope, though it may cause mild pacing-related symptoms.
May aid diagnosis in unexplained syncope; leaves open validation in larger prospective studies.
Unexplained syncope is the main indication for the implantation of a diagnostic pacemaker. Studies on those implanted have shown that in patients with unexplained syncope, the diagnosis of paroxysmal bradycardia was feasible and reliable. The present study was designed to evaluate a new bradycardia diagnosis algorithm, loaded in a dual chamber pacemaker, in 24 patients considered as candidates for diagnostic pacemakers. During a mean follow-up of 153 days, at least one bradycardia episode was recorded in 13 patients. The mean number of detected bradycardias was 6 and the median was 3. The mean delay between the algorithm activation and the first bradycardia episode was 67 days. The mechanism of bradycardia was atrioventricular block in 6 patients, sinus node dysfunction in 6 patients, and consecutive blocked atrial premature beats in 1 patient, as indicated by the event markers. In 11 patients bradycardia was recorded during the daytime only or day and night. In two patients the episodes were recorded only at night. Overall, the algorithm was well-tolerated; however, some mild symptoms were observed due to the method of bradycardia determination, allowing bradycardia. Three patients were symptomatic as a direct result of the algorithm operation, and four patients had symptoms related to the single chamber operation of the pacemaker while functioning in the diagnosis mode (VDI). These symptoms were relieved with DDD pacing.
No takes yet. Share an insight, caveat, or question.
Lascault et al. (1995) conducted an observational in Unexplained syncope (n=24). Dual chamber pacemaker with a new bradycardia diagnosis algorithm was evaluated on Detection of at least one bradycardia episode. A new bradycardia diagnosis algorithm in a dual chamber pacemaker detected at least one bradycardia episode in 54.2% (13 of 24) of patients over a mean follow-up of 153 days.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: