Key result
Insertion of a microporous coated Austin-Moore type hip prosthesis in 34 patients resulted in 4 reoperations and 10 cases of radiolucent lines over a 3-year period.
Why the study?
Does a microporous coated Austin-Moore type prosthesis result in successful bone ingrowth and clinical stability in patients requiring hip replacement?
Observational (n=34)
Does a microporous coated Austin-Moore type prosthesis result in successful bone ingrowth and clinical stability in patients requiring hip replacement?
Early clinical evaluation of a microporous coated metal hip prosthesis revealed significant failure of bone ingrowth, highlighting the need for modifications in prosthesis design and postoperative management.
Answers are grounded in this paper and cite their sources.
High reoperation rates suggest ingrowth failure with this design; hypothesis-generating for coating modifications before further use.
In an attempt to find a better method of attaching a prosthesis to bone, a microporous coating system, with a pore width of approximately 50 microns was chosen for evaluation. A microporous coated Austin-Moore type prosthesis was inserted into 34 patients, and the results were studied for a three-year period. Four cases required reoperations, and ten cases demonstrated radiolucent lines around the prosthesis suggesting that bone ingrowth had not occurred. This article analyzes the reasons for failure of ingrowth and suggests modifications in the prosthesis design and postoperative management that hopefully will be effective.
No takes yet. Share an insight, caveat, or question.
Hugh U. Cameron (1982) conducted an observational in Need for hip prosthesis (n=34). Microporous coated Austin-Moore type prosthesis was evaluated on Reoperations. Insertion of a microporous coated Austin-Moore type hip prosthesis in 34 patients resulted in 4 reoperations and 10 cases of radiolucent lines over a 3-year period.
Synapse has enriched 10 related papers. Consider them for comparative context: