Answers are grounded in this paper and cite their sources.
Capillary zone electrophoresis with amperometric detection was employed for the determination of lincomycin and lincomycin B in bulk drug and pharmaceutical formulations. The electrode used was a 300μm diameter Cu disk electrode operated in a wall-jet configuration. With an applied potential of +675mV (vs Ag/AgCl) and 80cm capillary(i.d. 25μm) at 26kV in the electrophoretic medium of 0.lOmol/L NaOH, lincomycin and lincomycin B could be completely separated within 15 minutes with resolution of 2.3. The response was linear over the concentration of 10.0–150.0μg/mL with a correlation coefficient of 0.9998, and the detection limit of 1.0μg/mL for lincomycin. The results of analysis were 98.2–99.3% of lincomycin label claims, and compared favourably with those obtained by the official method. The method is simple, rapid and precise for routine analysis of underivatized lincomycin and lincomycim B in bulk drug and pharmaceutical formulations.
No takes yet. Share an insight, caveat, or question.
Jiannong Ye (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: