Answers are grounded in this paper and cite their sources.
Tissues from around surgical implants have been examined by optical microscopy and electron microscopy and by electron diffraction and x‐ray spectrographic techniques. The presence of small alloy fragments and carbides worn from cobalt‐chromium total joint replacements was confirmed. These foreign bodies had been phagocytosed by macrophages and there were no indications that the tissues were damaged by their presence. In 13 of 44 specimens there were granulomatous foci related apparently to the presence of numerous minute particles of unknown composition which had elicited vigorous macrophage reactions. In tissues near stainless steel joints too there were particles of the naked alloy taken up by macrophages and again the fibrous tissue appeared unaffected. Typically the tissues near to stainless steel implants are affected by cytosiderosis. It is deduced that the iron released by corrosion from stainless steel is processed by macrophages locally and there is formed a mixture of hydrated ferric oxides similar to hemosiderin. This may be looked upon as a detoxication mechanism. There is evidence that in some specimens the limit of tissue tolerance for stored iron has been exceeded locally with consequent pathological changes. Another common deposit around stainless steel implants occurs in large plaques. It attracts giant cells and there is evidence that it is fibrotic and cytotoxic.
No takes yet. Share an insight, caveat, or question.
George D. Winter (1974) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: