Answers are grounded in this paper and cite their sources.
A comprehensive, interchangeable, low-contact-stress, mobile-bearing knee prosthesis system was developed and used over a 12-year period with both cemented and cementless fixation. Individual components of the system included a bicruciate-retaining meniscal bearing, a posterior cruciate-retaining meniscal bearing, and a cruciate-sacrificing, rotating-platform tibial component mated to the same femoral and rotating patellar components. Survivorship analysis of each implant type was performed to identify specific failure modes and trends for long-term survival of the implants in a wide variety of primary knee arthroplasties. Clinically, there were 46 prostheses of the bicruciate type followed for up to 12 years, 57 prostheses of the posterior-cruciate type followed for up to six years, and 108 prostheses of the rotating-platform type followed for up to ten years. All knees in this study had rotating-bearing patellar prostheses. Cumulative survivorship analysis using an end point of implant revision or a poor knee score revealed a small early failure rate of each implant in the first three years, associated with technical positioning or undersizing errors. This study indicates a predictable long-term survival of both cruciate-retaining and cruciate-sacrificing mobile-bearing knee prostheses as well as rotating-bearing patellar prostheses when used in primary knee arthroplasties that minimized technical errors of insertion.
No takes yet. Share an insight, caveat, or question.
Buechel et al. (1990) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: