Answers are grounded in this paper and cite their sources.
A method has been developed to reduce a magnetic resonance image artifact arising from the Fourier transformation of data truncated in the phase-encoding direction. The discontinuity is eliminated between the phase of the complex data sample associated with the maximum negative phase-encoding gradient and the phase of the maximum positive gradient sample. Images retain sharper edges than if low-pass filtering is employed. The method is demonstrated on clinical images, then an explanation of the method is presented.
No takes yet. Share an insight, caveat, or question.
Wood et al. (1985) studied this question.