Key result
Dynamic flow type testing resulted in higher calcium deposition on porcine aortic valve leaflets compared to conventional batch testing after 14 days.
A novel in vitro dynamic flow type calcification testing method allows for more rapid and pronounced assessment of bioprosthetic heart valve calcification compared to conventional batch testing.
Answers are grounded in this paper and cite their sources.
May accelerate in vitro screening of bioprosthetic valve durability; leaves open correlation with in vivo outcomes.
A dynamic flow type testing to study calcification was self-designed to investigate calcification in bioprosthetic heart valves. The apparatus consists of a container into which leaflets from a porcine aortic valve are placed, a chamber that contains calcium solution, and a peristaltic pump that provides a continuous supply of the solution toward the container. Efficacy of the apparatus was compared with the conventional batch type calcification testing at 37 degrees C through measuring the amount of calcium and phosphate deposited by inductively coupled plasma (ICP) and scanning electron microscope (SEM). After 14 days, calcium levels detected from the calcified deposit on leaflets were 470.4 +/- 37.0 microg/cm3 in the flow type testing whereas in the batch type testing levels were 81.0 +/- 6.7 microg/cm3. Though the calcium level on the leaflet increased as the exposure time to calcium solution increased in both testings, the rate and the tendency of calcification could be assessed very rapidly by flow type testing in comparison with batch type testing. [Ca]/[P] molar ratio decreased over time, and after 14 days, the ratio was close to 1.83 +/- 0.18 in the flow type testing. The ratio could not be determined in the batch type testing because the deposit was too small to assess. The descending rate of [Ca]/[P] molar ratio demonstrates that deposited calcium-complex at the earliest stage may interact with inorganic phosphate ions to create a calcified deposit mineral precursor. This in vitro dynamic flow type calcification testing was a favorable tool for rapid investigation of calcification.
No takes yet. Share an insight, caveat, or question.
Park et al. (2000) studied Calcification in bioprosthetic heart valves. Dynamic flow type testing vs. Conventional batch type calcification testing was evaluated on Calcium levels detected from the calcified deposit on leaflets. Dynamic flow type testing resulted in higher calcium deposition on porcine aortic valve leaflets compared to conventional batch testing after 14 days.
Synapse has enriched 10 related papers. Consider them for comparative context: