A method for the determination of nonionics in low concn (0舑20 ppm) has been developed. This method is based on the formation of a blue complex between ammonium cobaltothiocyanate reagent and a polyethoxylated compound. This complex is extracted into benzene from a saturated salt solution and measured with a spectrophotometer at 320 mƁ. The absorbance reading is compared to a standard. The method requires only a single extraction and no preliminary concn step is necessary. It is sensitive to polyethoxylated compounds containing three or more moles of ethylene oxide. We have used this method to detect as little as 0.01 mg of nonionic in 100 ml of sample (0.1 ppm). Reaction conditions, sensitivity differences between compounds, interferences and suitability for use in biodegradation studies have been investigated. The accuracy and repeatability of this method is ŷ3% relative at 20 ppm concn.