The effect on the shape of “metastable peaks” arising from decompositions throughout the whole flight path in a mass spectrometer is considered. The effects of discriminations against product ions are calculated; the effects of finite beam dimensions, angular spreading and kinetic energy of separation of fragments during the decomposition are discussed theoretically and illustrated by practical examples.
No takes yet. Share an insight, caveat, or question.
Beynon et al. (1967) studied this question.